Conductivity Probe

  • Range of Conductivity Probe
    • 0 to 2000 µS/cm (0 to 1000 mg/L TDS)
  • Total System Accuracy using Factory Calibration
    • ±3% of full-scale reading
  • Accuracy using custom calibration
    • ±1% of full-scale reading
  • Response time
    • 90% of full-scale reading in 5 seconds
    • 100% of full-scale in 20 seconds
  • Temperature compensation
    • PT100
    • Typical Impedance Error with Temperature (30ppm/°C)
  • Cell constant
    • 1.03 cm-1
  • Description
    • ABS body, parallel titanium stainless steel electrodes
  • Dimensions
    • 12 mm OD and 140 mm length

The Mantis Conductivity Probe can be used to measure conductivity in many of your science projects.  The probe is very robust but a few situations should be avoided with the conductivity probe:

  • NEVER submerge the plastic sensor end (the corded end) in any liquid as it is not waterproof
  • NEVER exceed the 0 – 75⁰ C maximum temperature or you may permanently damage the probe

Some things you should do:

  • ALWAYS wash the conductivity probe with distilled water and dry the sensor after ever use
  • ALWAYS keep the plastic end of the sensor dry
  • ALWAYS follow the storage recommendations for the conductivity sensor and store in a dry place

The Conductivity Probe is ideal for experiments in the chemistry lab, biology lab or physics lab.  The probe is also ideal in gathering data outside of a lab situation!

  • Measuring Total Dissolved Solids in a lake or river or a lab sample
  • Measuring rates of chemical reactions in which dissolved ions and solution conductivity changes with time
  • Measuring total concentration of ions in a solution
  • Measuring the difference between ionic and molecular components in a liquid solution
  • Material Property Analysis
  • Electrochemical Analysis
  • Bioelectrical Impedance Analysis